What is Nerve Mapping and how does it work?

Nerve mapping is a technique whereby superficial peripheral nerves can be traced and located percutaneously. The technique enables the anesthesiologist to determine the site for needle insertion prior to puncturing the skin.


If a neuromuscular response is not achieved for the specific Nerve Mapping Probe position should one rather increase the Current Amplitude, or the Pulse Width?

There is an optimal Pulse Width at which a specific nerve is most excitable. This is called the chronaxie threshold. It is preferable to keep the Pulse Width as close to this value for the related nerve or nerve plexus, as the Nerve Stimulator allows, then increase only the current.

 


What are the advantages of using Nerve Mapping?

Due to the fact that one pre-determines the surface location of the nerve and thus the optimal entry point for the needle, the technique reduces the need for multiple needle insertions and discomfort to the patient.


For which Regional Blocking procedures are Nerve Mapping appropriate.

The nerve mapping technique may be used for various approaches to the brachial plexus, as well as the axillary, musculo-cutaneous, ulnar, median and radial nerve blocks of the upper limb; and the femoral, sciatic and popliteal nerve blocks in the lower limb. Surface nerve mapping is particular useful where classic anatomical landmarks are absent or difficult to define, for example in children with contractures (arthrogryposis multiplex congenital; burns) or with major congenital limb defects.


What is Coulomb’s Law and how does it apply to Nerve Mapping.

It relates the charge intensity necessary to elicit a neuromuscular response to the distance between the probe and the nerve.


How is the STIMPOD unique with regards to implementing the Nerve Mapping Technique?

The STIMPOD allows the user to use the needle and probe together, eliminating the need to change cables or make markings on the patient.


Can an approximation be made with regards to current strength versus distance of the nerve?

Yes, the following approximation can be made concerning depth penetration. (Pulse width of 200us and more in patients of normal weight)
  • 5mA = 1-2cm depth penetration
  • 20mA = 3-4cm depth penetration

As shown in the figure below, the current density radiates outward from the nerve mapping pen in a spherical form. It also shows the decrease in behavioral density as the distance is increased from the source. Axons with a larger diameter exhibit a lower activation threshold than small axons. This results in electrical stimulation activating larger axons first before activating the smaller axons. Looking at the behavioral density in figure shown, most of the axons will be activated close to the probe, whereas only the larger diameter axons will be activated further away from the probe.

Graph01_01.png

There are two options to increase the energy delivered to a nerve without changing the distance from the electrode – increase the current amplitude and/or increase the pulse width.


It should be noted that, although there are many published values for chronaxie for various excitable tissues, the range of variability for a given tissue type is quite large. It is generally assumed, however, that nerves can be classified according to their chronaxie thresholds as follows:

Classification

Chronaxie

Sensory functions

A(alpha)

40-100 us

Predominantly motor neurons they also have the following sensory functions: Proprioception, hair receptors, vibratory sensors, high discrimination touch

A(delta)

150 us

Deep pressure and touch, pricking pain, cold

C

400 us

Crude touch and pressure, tickle, aching pain, cold, warmth

From the above table it would seem reasonable to deduce that the ideal pulse width to facilitate a motor nerve response (A alpha), would be around 100us. If one sets the nerve stimulator at 100us and increase the amplitude to 5mA giving a total charge of 500nC one would not get the same muscle response as if the setting is at 500us and 1mA, also giving a total Charge of 500nC. In the second case even though the total charge transferred to the nerve is the same, because of the chronaxie threshold of 100ms for the nerve, much of the energy transferred to the nerve after the 100us is wasted on the nerve.

Strength_Duration_Curve.png

This is clearly shown by the graph above. The strength duration curve (green) indicates the current necessary at the different pulse widths to facilitate a contraction. The energy cost or total charge is shown by the blue curve. It can be seen that the stimulation is most energy efficient at the chronaxie pulse of +- 80us width as would be expected. It should be noted how the energy cost increases when pulse width.

As a preference, keep the device at a 100us pulse width and adjust the current. If the device is already set at 20mA and the Nerve Mapping Probe does not elicit any neuromuscular response, increasing the pulse width to 300us will offer 3 x more charge, however bear in mind that the net effect on the nerve will not constitute a contraction which is 3 times more powerful.

 


 


 


In order for nerve excitation to take place, the energy delivered to the nerve must be high enough to trigger the threshold voltage. The current-distance relationship is governed by Coulomb’s Law:

E = K(Q/r2)
where E is the energy required, K is a constant, Q is the minimum current and r is the distance away from the electrode.

The equation shows what effect the distance has on the energy delivered.
2 x Distance = ¼ Energy

The equation shows what effect the distance has on the energy delivered.
2 x Distance = ¼ Energy

The equation also shows how you need to increase the current as your distance increases.
2 x Distance = 4 x Current


 The practical application of nerve mapping as a technique has always been a cumbersome technique. In order to use nerve mapping effectively, the position and angle of the probe must be recorded exactly before inserting the needle. Marking the probe position with a marker has proved unsuccessful and cumbersome, as this does not capture the exact position and angle relating to subcutaneous structures - i.e. bone structure, muscles etc. It is important to have the ability to keep the probe in its original position when inserting the needle. This will ensure exact positioning and angle of the needle. The STIMPOD facilitates this procedure by providing a combined Nerve Mapping/Locating cable. The unit will automatically switch between the probe and the needle, depending on which device is in contact with the skin. The STIMPOD will guide you through the entire procedure, switching between the probe and the needle as needed whilst keeping an eye out for high impedance and nerve proximity.



These are only approximations as the impedance structures of the surrounding tissue are very complex and differ from area to area. Obese patients can render the nerve mapping technique useless if your nerve stimulator does not have the option to significantly increase the current due to higher than normal impedance levels in the surrounding tissue.

The tip design also influences the current density, and as a result will influence the depth penetration of the probe. The figure illustrates the respective depth penetration from a large tip and a smaller tip. The current density is defined as the current applied over a specific area. The area is directly related to the radius of the tip and the part of the tip that is in contact with the skin. In the case of the larger tip, this area is significantly more than with the smaller tip. Because current density decreases as the distance from the source increases, the smaller tip will be able to activate far more axons than the larger tip at the same depth beneath the skin.