What is the purpose of the waveform display?

A perfect square wave indicates that the circuit impedance is acceptable. A non-square wave indicates that the circuit impedance is too high.


Why not just display a value for the impedance measurement rather than a graphic display of the waveform shape?

An impedance measurement offers a very inconsistent feedback, because of the varied impedance spectra of different types of tissue. The movement of the needle also adds to the continued variance in impedance.


What is Chronaxie and how does it affect the choice of pulse width?

Rheobase is the lowest current that will elicit a neuromuscular response for an infinitely long-duration stimulus. Chronaxie, the excitability constant, is the pulse width which will elicit a neuromuscular response for a stimulus current of twice rheobasic strength.


What does it mean if the waveform is only partially square and how does it relate to the stimulated muscle response?

It means that the maximum Voltage that the device can deliver is not enough to accommodate for the higher impedance encountered by the circuit. Depending on the pulse width of the square section of the waveform, the contraction from the stimulated muscle, may or may not represent an artefact.

A Perfectly square waveform indicates to the user that the device accommodates sufficiently for the impedance of the closed circuit path. This includes the connection between the ECG pad and the patient, as well as the probe or needle and patient. If the shape of the waveform does not resemble a perfect square wave, it means that the impedance of the circuit is higher than what the device can accommodate for, resulting in less current (charge) being delivered than what the current (charge) setting indicates.

A circuit impedance which is too high is usually caused by an old or dry ECG electrode. Make sure that the ECG electrode is in good condition. Ensure that the patient’s skin is not excessively dry where the ECG electrode is connected.

When using the nerve mapping probe, it is possible that the probe – skin connection impedance is too high. Ensure that the skin is moist enough in the area where the nerve mapping is intended.


The impedance of a circuit element is defined as the ratio of the phasor voltage across the element to the phasor current through the element:

In other words, impedance is the frequency dependant resistance as presented by its real and imaginary components. In the human body different tissues have different impedance spectra and can be modeled by an equivalent circuit comprising resistive, capacitive and inductive components.

When considering different effects of electrical stimulation, equivalent models of electrode and tissue impedance can help to explain certain stimulation phenomena which may seem inconsistent to the clinician.

The figure below shows an equivalent model, demonstrating the different impedances to be considered when analyzing the electrode, gel, epidermis, and dermis

Graph03_01.png

It should become clear that the total impedance path through which the electric field, generated between two electrodes, propagates is the complex result of the summation of the unique impedance spectra of various types of tissue. Each different stimulating pulse width, as well as its rise and fall times presents different frequency components which will result in different unique impedances.


It should be noted that, although there are many published values for chronaxie for various excitable tissues, the range of variability for a given tissue type is quite large. It is generally assumed, however, that nerves can be classified according to their chronaxie thresholds. 

From the table below it would seem reasonable to deduce that the ideal pulse width to facilitate a motor nerve response (A alpha), would be around 100us. If one sets the nerve stimulator at 100us and increase the amplitude to 5mA giving a total charge of 500nC one would not get the same muscle response as if the setting is at 500us and 1mA, also giving a total Charge of 500nC. In the second case even though the total charge transferred to the nerve is the same, because of the chronaxie threshold of 100ms for the nerve, much of the energy transferred to the nerve after the 100ms is wasted on the nerve.

Classification  Chronaxie  Sensory Functions 
 A (alpha) 40-100us  Predominantly motor neurons they also have the following sensory functions: Proprioception, hair receptors, vibratory sensors, high discrimination touch 
 A (delta) 150us  Deep pressure and touch, pricking pain, cold 
 C 400us   Crude touch and pressure, tickle, aching pain, cold, warmth

 

 

 


When one stimulates with a good current source, the shape and amplitude of your stimulus pulse will always be as selected, as long as the stimulator can deliver the voltage required to accommodate for the varying circuit impedance. All brands of stimulators are limited in the way they can accommodate varying impedances by their maximum voltage.

Graph03_02.png

The figure above shows a typical current and voltage stimulation response. V (channel 2) is measured across the two electrodes connected to a subject’s body. I (channel 1) is measured over a 10ohm resistor connected in series with one of the electrodes. The maximum current as displayed in this picture is 5mA. The maximum voltage necessary to facilitate this is approximately 40V. Even though stimulation was done with a 5mA, 1ms square wave stimulus, the approximately 80us negative current component is indicative of the reactive impedance of the combined electrode, tissue impedance.

Graph03_04.png

The figure above shows the stimulator at the same settings, however the impedance of the electrode/ epidermis interface was increased to a level where the device cannot supply enough voltage to facilitate the increased impedance.

It is clear that after approximately 140us the device could not deliver the required voltage. The current immediately dropped to around 4mA.

According to the discussion on chronaxie thresholds, though, it is quite likely thou that the second waveform will elicit a very similar response to the first waveform. This is due to the fact that the second waveform is ‘square’ for the first 140us  while the chronaxie of the nerve is 100us. this means that the drop in current (charge) supplied after 140us would have a limited effect on the nerve due o the fact that it would’ve been ‘wasted’ in any case.

One approach to offering the user an indication of the expected net stimulus effect would be to average out the total current delivered. This would give the user the impression that the observed response was equivalent to a perfect square wave of 1ms pulse width and 3.7mA amplitude. However due to the discussion in the paragraph above, it should be noted that most of the neuromuscular stimulating response was most likely facilitated in the first 100us of the stimulus at 5mA. It could then be misleading to simply look at the average stimulating amplitude.

In other words it could be argued that the stimulation as indicated in in the first figure and the one in the second figure elicit a similar neuromuscular stimulating response (contraction) with the electrodes positioned at exactly the same distance from the target nerve. If one then relied on the information presented of an actual average current transferred, one would have the erroneous impression that the cathode in eliciting a response in the case of the second figure would be closer to the nerve than the cathode that elicited the same response in the case of the first figure.

Bookmark and Share